Installation | |
Ambient temperature | 5 … 38 °C, with cabinet heating ± 0 °C |
Site | Indoor/outdoor installation (configurable) |
Gas inlet conditions | |
Sample gas pressure | - With pump, depressurized suction mode, selectable with internal or external pump
- Provision must be made for a pressure reduction for pressures greater than 1 200 mbar absolute
|
Pump performance | Adjustable to 60 … 80 Nl/h |
Sample gas temperature | Max. 45 °C, with moisture saturation |
Power supply | |
Supply 1 | 200 … 240 V AC, 47 … 63 Hz |
Supply 2 | 100 … 120 V AC, 47 … 63 Hz |
Power consumption | Approx. 180 VA (without cooler and sample preparation) |
Connection systems | |
Teflon hose | With PVDF screwed glands |
Connection systems | Metric (6 mm) or imperial (1/4") selectable |
Dimensions | |
Set BGA measuring system (W x H x D) | 600 x 781 x 600 mm |
Sample preparation (W x H x D) | 600 x 600 x 220 mm |
Weight | |
Set BGA measuring system | Approx. 50 kg |
Sample preparation | Approx. 22 kg |
System design | |
System housing | 3-part sheet-steel housing with window |
Degree of protection | IP54 |
Cabinet conditioning | Fan |
Cooling system | Peltier cyclone cooler (optional) |
Sample preparation | Max. six sample streams can be controlled using Logo module with fast loop pump in separate housing |
Analog outputs | Per component 0/2/4 … 20 mA; NAMUR, floating, max. load 750 ? |
Measured components / measuring ranges | |
CH4 | 0 … 100 vol.% to 0 … 20 vol.% (NDIR) |
CO2 | 0 … 100 vol.% to 0 … 20 vol.% (NDIR) |
O2 | 0 … 25 vol.% to 0 … 5 vol.% (electrochemical or paramagnetic optionally selectable) |
H2S | 0 … 5 ppm to 0 … 50 ppm (electrochemical); optional |
Safety assemblies | |
Assembly 1 | Detonation protection F501 |
Assembly 2 | Flow measurement with limit monitoring at the output |
Assembly 3 | LEL monitoring (optional) |
Comment | - The system concept of the Set BGA is based on the preconfigured ULTRAMAT 23 solutions (7MB2335-…, 7MB2337-…)
- The technical performance data concerning the measuring response correspond to the catalog data of the ULTRAMAT 23. The preconfigured version does not contain any ULTRAMAT 23 add-ons or retrofitting sets.
|