Cross-duct sensor CD 6General information | | Design | Transmitter and detector units, connected by a sensor cable | Materials | Stainless steel (1.4305/303), aluminum | Installation | Vertical or parallel to the gas flow | Laser protection class | Class 1, safe to the eye | Explosion protection | II 1 G Ex ia op to IIC T4 Ga II 1 D Ex ia op to IIIC T135 °C Da A defined leak rate can only be guaranteed when using high-pressure window flanges. Otherwise, it may be necessary for the owner to carry out an evaluation in accordance with ATEX DEMKO 06 ATEX 139648X; IECEx UL 13.0029X | Design, enclosure | | Degree of protection | IP65 | Dimensions | Diameter: 163, L: 450 mm | Purging gas tube in mm | 400 (370 net) x 44 x 40 800 (770 net) x 54 x 40 1 200 (1 170 net) x 54 x 40 | Weight | 2 x approx. 11 kg | Mounting | DN 65/PN 6, DN 80/PN 16 or ANSI 4"/150 lbs | Please note: - For purging tubes with a length of 800 and 1 200 mm, the wall thickness must not exceed 200 mm with DN 65/PN 6 connections. To carry out measurements with thicker walls, please contact Siemens.
- The optimum adjustment of the flanges can change with high differences in temperature between the process and environment depending on the type of assembly.
| | Electrical characteristics | | Power supply | 24 V DC, supply from central unit via hybrid cable | Power consumption | < 2 W with non-Ex configuration, max. 0.6 W with Ex configuration | Climatic conditions | | Sensor temperature | | Non-Ex | -20 ... +70 °C in operation -30 ... +70 °C during transport and storage | Ex | -20 ... +60 °C in operation -30 ... +70 °C during transport and storage | Humidity | < 95 % RH, above dew point | Pressure | 800 ... 1 100 hPa | Temperature range on the sensor side of the process interface (connection plate) | -20 ... +70 °C | Measuring conditions | | Measurement path | 0.3 ... 12 m (other path lengths on request) | Dust load | The influence of dust is very complex and depends on the path length and particle size. The optical attenuation increases exponentially at longer path lengths. Smaller particles also have a very large influence on the optical attenuation. With high dust load, long path length and small particle size, the technical support at Siemens should be consulted. |
AccessoriesPurging | | Nitrogen is permissible as the purging gas for the sensor side. Nitrogen, steam, air and gases which are not subject to the pressure equipment directive Cat. 2 are permissible as purging gases for the process side. | | Purging with instrument air, N2 | | - Max. overpressure in the sensor
| < 500 hPa | | | | According to ISO 8573-1:2010 [2:3:3] Note: It is sufficient if the pressure condensation point is min. 10 K below the minimum ambient temperature. | | Purity better than 99.7 %. For oxygen measurements, an O2 content < 0.01% in the purging gas. Optical path length ≥ 1 m, min. 5% oxygen in the process gas. | - Maximum flow rate (process purging)
| 500 l/min | | Benchmark: < -10 °C, condensation on the optics must be avoided | Blower purging | | | 40 hPa | | 850 l/min | | 370 W | - Degree of protection (fan)
| IP54, cover required to protect against rain | Steam purging | | | Overheated | | 240 °C | | > 4 000 hPa | | 16 000 hPa, refers to a volume flow of approx. 1 100 l/min |
Hybrid and sensor cablesGeneral information | | Configuration hybrid cable | Two optical fibers and two twisted copper wires in one cable for 24 V DC. Single-mode optical fiber fabricated at both ends with E2000 angle connectors. Multimode optical fiber configured at both ends with SMA connectors. Cable is flame-retardant, very good resistance to oil, gasoline, acids and alkalis, outer sheath UV-resistant | Cable sheath | Oil-resistant polyurethane | Dimensions | - An external power supply must be additionally ordered for > 500 m
- For installation in hazardous zones, non-intrinsically-safe cables have to be spatially separated from intrinsically-safe lines
| | < 8.5 mm | | - Use in non-hazardous and Ex Zone 2: Up to 700 m
- Use in Ex Zone 0 and Zone 1: Up to 250 m
| Weight | 75 kg/km | Maximum tensile force | 200 N | Maximum lateral pressure | 1 000 N/cm | Impact resistance | 200 N/cm | Maximum tensile strength | 500 N | Minimum bending radius | 12 cm | Climatic conditions | | Ambient temperature | -40 ... +70 °C during transport, storage and operation -5 ... +50 °C during cable installation | Humidity | < 95% rel. humidity, above dew point (in operation and storage) |
|
|