Cross-duct sensor CD 6
|
General information
|
|
Design
|
Transmitter and detector units, connected by a sensor cable
|
|
Materials
|
Stainless steel, aluminum
|
|
Installation
|
Vertical or parallel to the gas flow
|
|
Laser protection class
|
Class 1, safe to the eye
|
|
Explosion protection
|
II 1 G Ex ia op to IIC T4 Ga II 1 D Ex ia op to IIIC T135 °C Da
A defined leak rate can only be guaranteed when using high-pressure window flanges. Otherwise, it may be necessary for the owner to carry out an evaluation in accordance with ATEX DEMKO 06 ATEX 139648X; IECEx UL 13.0029X
|
|
Design, housing
|
|
Degree of protection
|
IP65
|
|
Dimensions
|
Diameter: 163, L: 450 mm
|
|
Purging gas tube in mm
|
400 (370 net) x 44 x 40
800 (770 net) x 44 x 40
1 200 (1 170 net) x 44 x 40
|
|
Weight
|
2 x approx. 11 kg
|
|
Mounting
|
DN 65/PN 6, DN 80/PN 16 or ANSI 4"/150 lbs
|
|
Please note:
- For purging tubes with a length of 800 and 1 200 mm, the wall thickness must not exceed 200 mm with DN 65/PN 6 connections. To carry out measurements with thicker walls, please contact Siemens.
- The optimum adjustment of the flanges can change with high differences in temperature depending on the type of assembly.
|
|
Electrical characteristics
|
|
Power supply
|
24 V DC, supply from central unit via hybrid cable
|
|
Power consumption
|
< 2 W with non-Ex configuration, max. 0.6 W with Ex configuration
|
|
Climatic conditions
|
|
Sensor temperature
|
|
|
Non-Ex
|
-20 ... +70 °C during operation,
-30 ... +70 °C during storage and transportation
|
|
Haz:
|
-20 ... +60 °C during operation,
-30 ... +70 °C during storage and transportation
|
|
Humidity
|
< 95 % RH, above dew point
|
|
Pressure
|
800 ... 1 100 hPa
|
|
Temperature range on the sensor side of the process interface (connection plate)
|
-20 … +70 °C
|
|
Measuring conditions
|
|
Measurement path
|
0.3 ... 12 m (other lengths on request)
|
|
Dust load
|
The influence of dust is very complex and depends on the path length and particle size. The optical attenuation increases exponentially at longer path lengths. Smaller particles also have a large influence on the optical attenuation. With high dust load, long path length and small particle size, the technical support at Siemens should be consulted.
|
Accessories
|
Purging
|
|
Nitrogen is permissible as the purging gas for the sensor side. Nitrogen, steam, air and gases which are not subject to the pressure equipment directive Cat. 2 are permissible as purging gases for the process side.
|
|
Purging with instrument air, N2
|
|
- Max. overpressure in the sensor
|
< 500 hPa
|
|
|
|
|
|
According to ISO 8573-1:2010 [2:3:3]
Note: It is sufficient if the pressure condensation point is min. 10 K below the minimum ambient temperature.
|
|
|
Purity better than 99.7 %. For oxygen measurements, an O2 content < 0.01% in the purging gas (optical path length ? 1 m, min. 5% oxygen in the process gas)
|
- Maximum flow rate (process purging)
|
500 l/min
|
|
|
Benchmark: < -10 °C, condensation on the optics must be avoided
|
|
Blower purging
|
|
|
|
40 hPa
|
|
|
850 l/min
|
|
|
370 W
|
- Degree of protection (fan)
|
IP54, cover required to protect against rain
|
|
Steam purging
|
|
|
|
Overheated
|
|
|
240 °C
|
|
|
> 4 000 hPa
|
|
|
16 000 hPa, refers to a volume flow of approx. 1 100 l/min
|
Hybrid and sensor cables
|
General information
|
|
Configuration hybrid cable
|
Two optical fibers and two twisted copper wires in one cable for 24 V DC. Single-mode optical fiber configured at both ends with E2000 angle connectors. Multimode optical fiber configured at both ends with SMA connectors.
Cable is flame-retardant, very good resistance to oil, gasoline, acids and alkalis, outer sheath UV-resistant
|
|
Cable sheath
|
Oil-resistant polyurethane
|
|
Dimensions
|
- For > 500 m, an external power supply must be additionally ordered
- For installation in hazardous zones, non-intrinsically-safe cables have to be spatially separated from intrinsically-safe lines
|
|
|
< 8.5 mm
|
|
|
- Use in non-hazardous and Ex Zone 2: Up to 700 m
- Use in Ex Zone 0 and Zone 1: Up to 250 m
|
|
Weight
|
75 kg/km
|
|
Maximum tensile force
|
200 N
|
|
Maximum lateral pressure
|
1 000 N/cm
|
|
Impact resistance
|
200 N/cm
|
|
Maximum tensile strength
|
500 N
|
|
Minimum bending radius
|
12 cm
|
|
Climatic conditions
|
|
Ambient temperature
|
-40 ... +70 °C during transport, storage and operation
-5 ... +50°C during laying
|
|
Humidity
|
< 95% rel. humidity, above dew point (in operation and storage)
|
|
|