The flow measuring principle is based on the Coriolis effect. The flowmeter consists of a system FC410 or FC430 or a combination of a sensor type MASS 2100/FC300/FCS200 and a transmitter type MASS 6000/SIFLOW FC070/FCT010 and FCT030. The SITRANS F C sensors are energized by an electro-mechanical driver circuit which oscillates the pipe at its resonant frequency. Two pick-ups, 1 and 2 are placed symmetrically on both sides of the driver. When liquid or gas flows through the sensor, Coriolis force will act on the measuring pipe and cause a pipe deflection which can be measured as a phase shift on pick-up 1 and 2. The phase shift is proportional to the mass flow rate. The amplitude of the driver is automatically regulated to ensure a stable output from the 2 pick-ups. The temperature of the sensor is measured by a Pt1000. For MC2 the temperature is measured with a Pt1000. The flow-proportional signal from the 2 pick-ups, the temperature measurement and the driver frequency are fed into the SITRANS F C transmitter for calculations of mass, volume, fraction, temperature and density. The signal transfer function is based on a DFT technology (Discrete Fourier Transformation). The transmitter has a built-in noise filter, which can be used to improve the meter’s performance if the installation and application conditions are not ideal. Typically influence from process noise such as pump pulsations, mechanical vibrations, oscillating valves can be reduced considerably. SENSORPROM and SensorFlash flow memory units FC410 flow transmitters communicate via MODBUS RTU and FC430 via HART/Modbus/PROFIBUS DP/PROFIBUS PA. |