General information | |
Ambient temperature | -30 ... 55 °C (with heating) |
Explosion protection | The system Set CV is based on individually certified components according to ATEX 2G. A site permit issued by a regulatory authority is required (not included in scope of supply). |
Supply voltage | 230 V AC, 115 V AC or 24 V DC |
Max. permissible pressure at input of high-pressure reduction | 16 000 kPa |
Max. permissible pressure at input of sample preparation system with pressure regulator | 16 000 kPa |
Max. permissible pressure at output of sample preparation system | 10 … 60 kPa above atmospheric pressure Notice: Sample must not contain ethine! |
Sampling | |
Lance | Outer diameter 6 mm and inner diameter 2 mm or outer diameter 12 mm and inner diameter 8 mm Length 1 000 mm |
Special probe with BTU diaphragm and pressure reduction, non-retractable | - Immersion depth approx. 380 mm
- Protection pipe outer diameter 25.4 mm
|
Process connection | - At flange DN 65 PN 16 Form C, max. 1 600 kPa gas pressure
- At flange DN 65 PN 160 Form E, max. 16 000 kPa gas pressure
- At flange ANSI, 2-inch, 300 lbs RF
- At flange ANSI, 2-inch, 2 500 lbs RF
|
High-pressure reduction in casing with pressure regulators, optional heating | - Casing with dimensions 385 x 485 x 380 mm
- Primary pressure 16 000 kPa, output pressure 100/170 kPa (power consumption 150 W)
|
Pipe base | 2-inch pipe base for free-standing transmitter box, 1 700 mm high |
Heated line | - Heating power is 38 W/m
- Self-regulating heating up to approximately 80 °C
- Outer material is PE corrugated hose with 43 mm outer diameter
|
Sample preparation | For 1 to 3 sample gas flow, plus calibration gas flow |
Basic configuration | 0.5 µm filter, 3/2-way solenoid valve, flowmeter, overflow valve and stopcock |
Pressure regulator, optional heating | Primary pressure 16 000 kPa, output pressure 100/170 kPa (power consumption 150 W) |
Simple sample injection | One 3/2-way solenoid valve per sample gas stream |
DB&B sample injection | Two 3/2-way solenoid valves per sample gas flow |
On mounting plate | 682 x 482 x 3 mm |
In protective casing | 800 x 600 x 480 mm |
Protective casing heating | Power consumption 300 W adjustable between 10 and 40 °C, in increments of 5. |
Pipe base | 2-inch pipe base, 1 700 mm high |
Additional filters | Aerosol, glycol |
System components | |
On mounting plate | 650 x 422 x 3 mm |
In protective casing | 750 x 520 x 430 mm |
Protective casing heating | Power consumption 400 W adjustable between 10 and 40 °C, in increments of 5. |
Terminal box | 340 x 170 x 91 mm |
Pipe base | 2-inch pipe base, 1700 mm high |
Heated line | - Heating power is 38 W/m
- Self-regulating heating up to approximately 80 °C
- Outer material is PE corrugated hose with 43 mm outer diameter
|
Gas supply | |
Transfer station | Max. cylinder pressure 20 000 kPa Output pressure 50 ... 1 000 kPa |
Pipe base | 2-inch pipe base, mounted Height 2 200 mm |
Gas cylinder cabinet | Gas cylinder cabinet for two 50 l carrier gas cylinders and one 10 l calibration gas cylinder. Dimensions are 1 250 x 400 x 2 050 mm |
Cylinder heating | The cylinder heating system is dimensioned for 10 l calibration gas cylinders and ensures that no condensation occurs within the gas cylinder. |
Cylinder pressure reducer | Cylinder pressure reducer for reducing primary pressure of max. 30 000 kPa to back pressure 0 … 400 kPa |
Heated line | - Heating power is 38 W/m
- Self-regulating heating up to approximately 80 °C
- Outer material is PE corrugated hose with 43 mm outer diameter
|
Calibration gas | |
Gas mixture 1 ... 6 | The gas mixture is stable for 36 months. |