19" rack unit with 3 height units (HU) for installation
in hinged frames
in cabinets with or without telescopic rails
Gas connections for sample gas inlet and outlet: for pipe diameter 6 mm
Purging gas connections (optional), purging gas connection for 6 mm or 1/4" hose (optional)
Wall-mounted device
Gas connections for sample gas inlet and outlet: Pipe union for pipe diameter 6 mm (directly on the analyzer modules)
Purging gas connections: Pipe diameter 12 mm
Display and operator panel
LCD panel for simultaneous display of:
Measured value
Status line
Measuring ranges
Menu-driven operation for parameterization, test functions, adjustment
Operator support in plain text
Operating software (10 languages)
Display and operator panel of the SIPROCESS GA700 devices
Inputs and outputs
8 binary inputs, designed for 24 V, floating, freely configurable (e.g. for measurement range switchover, processing of external signals from sample preparation)
8 relay outputs, with changeover contacts, freely configurable (e.g. for faults, maintenance requests, limit alarms, external solenoid valves)
Ethernet connection contained in the base unit (connection on the rear side, Ethernet RJ 45, 100 MBit)
Service interface (front side); Ethernet RJ 45, 100 MBit.
Interface modules
Interface module 1.1: Adds 8 inputs and 12 outputs to the digital interfaces of the base unit.
Interface module 2.1: One analog output per measured component (max. 6, 0 to 20 mA, 4 to 20 mA or parameterizable according to NAMUR), plus 6 digital outputs
Interface Module 2.2: Three analog outputs and two analogue inputs per analyzer module, plus 4 digital inputs.