SITRANS FUT1010 | |
Input | |
Flow range (Gas) | ± 36.5 m/s (± 120 f/s) for DN 100 ... DN 200 (4" ... 8") pipes bi-directional ± 30.5 m/s (± 100 ft/s) for DN 250 ... DN 600 (10" ... 24") pipes bi-directional |
Flow range (Liquid) | ± 12 m/s (± 40 f/s) including zero flow, bi-directional |
Flow sensitivity | 0.0003 m/s (0.001 f/s) flow rate independent |
Flow temperature range | -28 ... +93 °C (-20 ... +200°F) |
Analog inputs | 4 x 4 ... 20 mA, (Programmable to Density, Pressure, viscosity or Temperature) |
Output | |
Standard outputs | - 4x isolated 4 ... 20 mA, programmable
- 2x 0 ... 10 V DC, programmable
- 4x Digital Pulse outputs (2x open collector and 2x 0-5V TTL) One each for positive flow, one each for negative flow
- Standard VT100 RS 232, Optional HART, BACnet MSTP/BACnet IP, Modbus RTU & TCP/IP, Ethernet IP, Johnson N2
|
Status/Alarm I/O | - Programmable, 4x Form C Relays
- Clear Switch Input Totalizer Hold Switch Input
|
Calibrated accuracy | |
Gas | |
| 0.5 ... 1.0 % (4” ... 6” < 0.15 %) |
| < 0.5 % |
| < 0.2 % |
Liquid | |
| 0.5 ... 1.0 % (4” ... 6” < 0.25 %) |
| < 0.5 % |
| < 0.15 % |
Repeatability | ± 0.05 ... 0.1 % |
Data refresh rate | 5 Hz |
Design | |
Design Flow transmitter | |
Dimensions | see SITRANS F US Clamp-on “System info and selection guide“ |
Weight | see diagrams |
Power supply | |
Power supply | 90 ... 240 V AC, 50 … 60 Hz, 30 VA or 9 ... 36 V DC, 12 W |
| | |
Indication and operation | |
Data logger memory | 1 MByte, programmable for all available data variables |
Display | 128 x 240 pixel LCD with backlight |
Keypad | 33 keypad buttons with tactile feedback |
Language options | English, Spanish, German, Italian, French |
Design Flow sensor | |
Nominal pipe sizes | 4" ... 24" (DN 100 ... DN 600) |
Pipe material specification | API 5L ERW |
Temperature tap | ?" |
Pressure tap | ?" |
Flange class | |
| 150, 300, 600 |
| 300, 600 |
Flange specification | - ASME B16.5
- Liquid 150, 300, 600
- Gas 300, 600
|
Flange facing | Raised face weld neck |
Flange material | A105 |
Flow sensor paths | Two, three, or four |
Sensor length | See diagram |
Design temperature | -28 ... +93 °C (-20 ... +200 °F) |
Exterior finish | Marine/offshore grade per ASTM B117 |
Optional pipe sections | - 10 D upstream (with optional flow conditioner)
- 5 D downstream
|
Certificates and approvals | |
Flow transmitter IP65 (NEMA 4X) | |
FM and CSA | N-I Class I, Div 2 S Class II, Div 2 I.S. Class I, II, Div 1 |
ATEX | Ex II (1) G [Ex ia] IIC Ex II 3 (1) G Ex nC [ia] IIC T5 |
CE markings | EMC 2004/108/EC ATEX 94/9/EC |
Flow Transmitter - IP66 (NEMA 7) | |
FM and CSA | Ex Class I, Div 1 D-I Class II, Div 1 N-I Class I, Div 2 S Class II, Div 2 I.S. Class I, II, Div 1 |
ATEX | Ex II (1) G [Ex ia] IIC Ex II 3 (1) G Ex nC [ia] IIC T5 Ex II 2 (1) G Ex d [ia IIC] IIB + H2 T5 |
CE markings | EMC 2004/108/EC ATEX 94/9/EC |
Sensor | |
FM and CSA | I.S. Class I, Div 1 N-I Class I, Div 2 S Class II, Div 2 |
ATEX | Ex II 1 G Ex ia IIC T5 |
CE markings | EMC 2004/108/EC PED 97/23/EEC ATEX 94/9/EC |