The modular measuring system consists of the following components:
Once it has passed through the heated cable, the sample gas flows into a two-stage compressor gas cooler. Between the 1st and 2nd cooler stage there is sample gas pump with integrated gas return for regulating the sample gas flows. Once it has passed through the sample gas cooler, the gas path splits into different partial lines to supply up to three analyzers simultaneously with sample gas. An additional partial flow lets the sample gas excess flow out over a bypass. To protect the analyzers, a condensate blocker is located directly upstream from the analyzers; it closes off the gas path when condensate enters the path. A three-way valve is installed upstream from the pump to supply the zero gas for automatic zero-point calibration. By default, the Set CEM CERT is operated by means of a touch screen panel (SIMATIC HMI, KTP700 BASIC) on the front of the measuring cabinet. Start menu on the SIMATIC HMI touch screen panel |