Pressure transmitter SITRANS LH300 (submersible sensor) |
Mode of operation | |
Measuring principle | Piezo-resistive |
Input | |
Measured variable | Hydrostatic level |
Measuring range | Max. permissible operating pressure |
- 0 ... 1 mH2O (0 ... 3 ftH2O)
| - 1.5 bar (21.8 psi) (corresponds to 15 mH2O (45 ftH2O))
|
- 0 ... 2 mH2O (0 ... 6 ftH2O)
| - 1.5 bar (21.8 psi) (corresponds to 15 mH2O (45 ftH2O))
|
- 0 ... 3 mH2O (0 ... 9 ftH2O)
| - 1.5 bar (21.8 psi) (corresponds to 15 mH2O (45 ftH2O))
|
- 0 ... 4 mH2O (0 ... 12 ftH2O)
| - 2 bar (29 psi) (corresponds to 20 mH2O (60 ftH2O))
|
- 0 ... 5 mH2O (0 ... 15 ftH2O)
| - 2 bar (29 psi) (corresponds to 20 mH2O (60 ftH2O))
|
- 0 ... 6 mH2O (0 ... 18 ftH2O)
| - 2 bar (29 psi) (corresponds to 20 mH2O (60 ftH2O))
|
- 0 ... 10 mH2O (0 ... 30 ftH2O)
| - 5.0 bar (72.5 psi) (corresponds to 50 mH2O (150 ftH2O))
|
- 0 ... 20 mH2O (0 ... 60 ftH2O)
| - 10.0 bar (145 psi) (corresponds to 100 mH2O (300 ftH2O))
|
- 0 ... 40 mH2O (0 ... 120 ftH2O)
| - 20.0 bar (290 psi) (corresponds to 200 mH2O (600 ftH2O))
|
Special measuring range | |
- Up to 100 mH2O (0 ... 300 ftH2O)
| - 20.0 bar (290 psi) (corresponds to 200 mH2O (600 ftH2O))
|
- Up to 160 mH2O (0 ... 480 ftH2O)
| - 24.0 bar (348 psi) (corresponds to 348 mH2O (720 ftH2O))
|
Measuring range | |
| |
| |
| |
| |
| |
| |
| |
| |
| |
Special measuring range | |
| |
| |
Output | |
Output signal | 4 ... 20 mA |
Measuring accuracy | According to IEC 60770-1 |
Error in measurement at limit setting including hysteresis and reproducibility | ? 0.15 % of full-scale value (typical) ? 0.3 % of full-scale value (maximum) |
Influence of ambient temperature | ? 0.05 %/10 K of full-scale value (zero and span) |
Long-term stability | ? 0.15 % of full-scale value/year (zero and span) |
Rated conditions | |
Ambient conditions | |
| -10 ... +80 °C (14 ... 176 °F) |
| -20 ... +80 °C (-4 ... +176 °F) |
Degree of protection according to IEC 60529 | IP68 |
Design | |
Weight | |
| ? 0.4 kg ( ? 0.88 lb) |
| 0.08 kg/m (? 0.059 lb/ft) |
Electrical connection | Cable with 2 conductors, vent pipe and integrated humidity filter |
Material | |
| Al2O3 ceramic, 99.6% |
| Stainless steel, mat. no. 1.4404/316L and 1.4539/904L (sea water applications) respectively |
| FPM (standard) EPDM (optional) |
| PE (standard/drinking water applications) FEP (for aggressive media) |
| Stainless steel, PPE or ETFE |
Auxiliary power | |
Terminal voltage on pressure transmitter UB | 10 ... 33 V DC 10 ... 30 V DC for transmitter with intrinsic safety explosion protection |
Certificates and approvals | |
Drinking water approval (ACS) | Pending |
Drinking water approval (WRAS) | Pending |
Drinking water approval (DVGW/KTW W270) | Pending |
EAC /EAC-Ex | № TC RU C-DE.ГБ05.В.00358 OC НАНИО «ЦСВЭ» |
Underwriters Laboratories (UL) | Pending |
Shipbuilding approval (LR) | Pending |
Shipbuilding approval (DNV/GL) | Pending |
Shipbuilding approval (BV) | Pending |
Shipbuilding approval (ABS) | Pending |
Pressure equipment directive | The device is not subject to the pressure equipment directive (PED 2014/68/EU) |
Explosion protection | |
IEC Ex ATEX EAC Ex | SEV 16 ATEX 0121 IEC Ex SEV 16.0003 |
| |
| II 1 G Ex ia IIC T4 Ga |