Measuring system D-FL 100
|
|
|
Measurements
|
- Flue gas velocity
- Volume flow 1)
|
|
Measuring ranges
|
- 2 ... 50 m/s
- 0 ... 3 000 000 m3/h
|
|
Measuring principle
|
Differential pressure
|
|
Sample gas temperature
|
Above dew point up to 450 °C, optional up to 850 °C
|
|
Sample gas pressure
|
-200 ... +200 hPa, optional higher
|
|
Duct diameter
|
0.4 ... 8 m
|
|
Ambient temperature
|
-20 ... +50 °C
|
|
Degree of protection
|
IP65
|
|
Measured value output
|
0/4 ... 20 mA, 500 ?
|
|
Digital outputs
|
2 relay outputs, permissible load 48 V/0.5 A
|
|
Digital inputs
|
None
|
|
Accuracy
|
< 2 % of the measuring range
|
|
Detection limit
|
< 3 m/s
|
|
Reference point drift
|
< 0.5 % of measuring range / month
|
|
Zero point drift
|
< 0.5% of measuring range / month
|
|
Auxiliary power
|
18 ... 32 VDC/1 A, 90 ... 264 VAC, 48 ... 62 Hz, 100 VA
|
|
Dimensions
|
|
|
|
380 x 160 x (300 + probe length) mm
|
|
|
0.4 ... 2 / 0.4 ... 4 / 0.4 ... 8 m
|
|
Weight
|
32 kg + 6.8 kg/m probe length
|
|
Purge-air supply (optional)
|
6 ... 8 bar for back purging
|