Flowmeter SITRANS FC410
|
|
|
Sizes
|
DN 15 (?") DN 25 (1") DN 50 (2") DN 80 (3")
|
|
Accuracy
|
± 0.10 %
|
|
Repeatability
|
± 0.05 %
|
|
Flow range (water @ 1 bar pressure loss)
|
DN 15: 3 700 kg/h (8 157 lb/h) DN 25: 11 500 kg/h (25 353 lb/h) DN 50: 52 000 kg/h (114 640 lb/h) DN 80: 136 000 kg/h (300 000 lb/h)
|
|
Architecture
|
Compact or remote configuration with selection of twelve languages including Chinese and Russian
|
|
Display
|
Full graphical display, 240 x 160 pixels
|
|
Power Supply
|
24 V DC ± 20 %; 110 mA
|
|
Weight
|
4.6 ... 50 kg
|
|
Materials
|
|
|
|
|
|
|
316L stainless steel or Hastelloy C22
|
|
|
304 stainless steel
|
|
|
Aluminum with corrosion-resistant coating
|
|
Enclosure rating (IP)
|
IP67
|
|
Pressure ratings
|
|
|
|
|
|
|
100 bar (1450 psi)
|
|
|
160 bar (2321 psi)
|
|
|
20 bar (DN15, DN25) 17 bar (DN50, DN80)
|
- Sensor enclosure burst pressure
|
> 160 bar (all sizes)
|
|
Temperature ratings
|
|
|
|
-50 ... 200 °C (-58 ... +392 °F)
|
|
|
-40 ... 60 °C (-40 ... +140 °F)
|
|
Process connections
|
|
|
|
EN 1092-1 B1, EN 1092-1 D, ANSI/ASME B16.5, JIS B 2220, DIN 11864-2
|
|
|
ASME B1.20 (NPT), ISO228-1 G (BSPP), VCO Quick-connect
|
|
|
DIN 11851, DIN 11864-1A, ISO 2853, SMS 1145
|
|
|
DIN 11864-3A, DIN 32676, ISO 2852
|
|
Approvals
|
|
|
|
ATEX, IECEx, FM, NEPSI, CSA, INMETRO (installed with flame-proof conduit)
|
|
|
PED, CRN
|
|
|
3A, EHEDG
|
|
|
Germanischer Lloyd/det Norske Veritas, Bureau Veritas, Lloyds of London, American Bureau of Shipping
|
|
NAMUR
|
NAMUR-compliant (e.g. NE 21, NE 41 and NE 132)
|
|
Communication
|
Modbus RTU
|
|
EMC performance
|
EN 61326-3-2
|
|
Mechanical load
|
18 to 400 Hz random The flow meter will mechanically tolerate 3.17 g RMS in all directions. Flow accuracy cannot be guaranteed under all conditions.
|