19" rack unit
|
General information
|
|
Operating position
|
Horizontal
|
|
Conformity
|
CE mark in accordance with EN 50081-1 and EN 50082-2
|
|
Design, enclosure
|
|
Weight without module
|
8.6 kg
|
|
Degree of protection
|
IP20 according to EN 60529
|
|
Electrical characteristics
|
|
Auxiliary power
|
100 ... 240 V AC (nominal range of use 85 ... 264 V), 50 ... 60 Hz (nominal range of use 47 ... 63 Hz)
|
|
Power consumption
|
Max. 280 VA
|
|
EMC interference immunity (electromagnetic compatibility)
|
In accordance with the standard requirements of NAMUR NE21 (05/2006) and EN 61326-1 (01/2008)
|
|
Electrical safety
|
In accordance with EN 61010-1, overvoltage category II
|
|
Electrical inputs and outputs
|
|
Relay outputs
|
8, with changeover contacts, can be freely parameterized, e.g. for measuring range identification; max. load: 24 V AC/DC/40 W (total load for all 8 relay outputs in continuous operation max. 160 W), potential-free, non-sparking
|
|
Binary inputs
|
8, designed for 24 V, potential-free, can be freely parameterized, e.g. for measurement range switchover
|
|
Ethernet interface (rear)
|
Ethernet RJ 45, 100 MBit
|
|
Service interface (front)
|
Ethernet RJ 45, 100 MBit
|
|
Option module 2.1
|
6 analog outputs, 0/4 to 20 mA, potential-free; maximum load 750 ? and 6 additional relay outputs, loading capacity: 24 V AC/DC/40 W, potential-free, non-sparking
|
|
Climatic conditions
|
|
Permissible operating altitude
|
3 000 m above sea level
|
|
Permissible ambient temperature (with one module; application-dependent with two modules)
|
- -30 ... +70°C during storage and transportation
- 0 ... 50°C during operation with one or two OXYMAT 7 analyzer modules
Ventilation slits must not be covered (recommended minimum upward clearance from the next device when installing 2 analyzer modules and at maximum ambient temperature: min. 1 HU)
|
|
Permissible humidity
|
< 90% RH (RH: relative humidity), during storage and transportation (dew point must not be undershot)
|
Wall housing
|
General information
|
|
|
Operating position
|
Vertical
|
|
Conformity
|
CE mark in accordance with EN 50081-1 and EN 50082-2
|
|
Design, enclosure
|
|
|
Weight without module
|
23 kg
|
|
Degree of protection
|
IP65 in accordance with EN 60529, restricted breathing enclosure to EN 50021
|
|
Electrical characteristics
|
|
|
Auxiliary power
|
100 ... 240 V AC (nominal range of use 85 ... 264 V), 50 ... 60 Hz (nominal range of use 47 ... 63 Hz)
|
|
Power consumption
|
Max. 280 VA
|
|
EMC interference immunity (electromagnetic compatibility)
|
In accordance with the standard requirements of NAMUR NE21 (05/2006) and EN 61326-1 (01/2008)
|
|
Electrical safety
|
In accordance with EN 61010-1, overvoltage category II
|
|
Gas inlet conditions
|
|
|
Purging gas pressure
|
|
|
|
< 100 hPa above atmospheric pressure
|
|
|
165 hPa above ambient pressure
|
|
Electrical inputs and outputs
|
|
|
Relay outputs
|
8, with changeover contacts, can be freely parameterized, e.g. for measuring range identification; max. load: 24 V AC/DC/40 W (total load for all 8 relay outputs in continuous operation max. 160 W), potential-free, non-sparking
|
|
Binary inputs
|
8, designed for 24 V, potential-free, can be freely parameterized, e.g. for measurement range switchover
|
|
Ethernet interface (bottom)
|
Ethernet RJ 45, 100 MBit
|
|
Service interface (bottom)
|
Ethernet RJ 45, 100 MBit
|
|
Option module 2.1
|
6 analog outputs, 0/4 to 20 mA, potential-free; maximum load 750 ? and 6 additional relay outputs, loading capacity: 24 V AC/DC/40 W, potential-free, non-sparking
|
|
Climatic conditions
|
|
|
Permissible operating altitude
|
3 000 m above sea level
|
|
Permissible ambient temperature (with one module; application-dependent with two modules)
|
- -30 ... +65? during storage and transportation
- 0 ... 50°C during operation with one or two OXYMAT 7 analyzer modules
|
|
Permissible humidity
|
< 90% RH (RH: relative humidity), during storage and transportation (dew point must not be undershot)
|
|
|