Possible applications
|
|
Container, conveyor, platform and roller table scales
|
|
Model
|
|
Ring-torsion load cell
|
|
Rated load/maximum load Emax.
|
|
60/130/280 kg
|
0.5/1/2/3.5/5/10 t
|
13/28/60 t
|
|
Accuracy class according to OIML R60
|
|
C31)
|
|
Max. load cell verification intervals nLC
|
|
3 000
|
|
Min. load cell verification intervals Vmin
|
|
E
max/16 000
|
E
max/17 500
|
|
Minimum application range Rmin(LC)
|
|
17 %
|
|
Combined error Fcomb
|
|
? ± 0.023 % Cn
|
|
Repeatability Fv
|
|
? ± 0.01 % Cn
|
|
Return of zero signal
|
|
? ± 0.0167 % Cn2)
|
|
Creep error Fcr
|
|
|
|
|
|
? ± 0.0245 % Cn2)
|
|
|
|
? ± 0.0053 % Cn2)
|
|
Temperature coefficient
|
|
|
|
|
|
? ± 0.004 % Cn/5K
|
|
|
|
? ± 0.004 % Cn/5K
|
|
Min. dead load Emin
|
|
? ± 0 % Emax
|
|
Safe load limit Lu
|
|
200 % Emax
|
150 % Emax
|
|
Ultimate load Ld
|
|
500 % Emax
|
300 % Emax
|
300 % Emax
|
|
Safe side load LIq
|
|
75 % Emax
|
100 % Emax
|
75 % Emax
|
|
Rated measuring path hn at Emax
|
|
0.07 mm
|
0.1 ± 0.02 mm
|
0.11 … 0.2 mm
|
|
Overload protection
|
|
Integrated
|
Integrated
|
Integrated at 13 t
|
|
Supply voltage Usr (reference value)
|
|
15 V
|
10 V
|
15 V
|
|
Supply voltage (range)
|
|
5 … 30 V+
|
|
Rated characteristic value Cn
|
|
1 mV/V
|
2 mV/V
|
2 mV/V
|
|
Tolerance Dc of characteristic value
|
|
Up to 500 kg: 0.01 mV/V from 500 kg: 0.1 mV/V
|
|
Tolerance Do of zero signal
|
|
? ± 1.0 % Cn
|
|
Input resistance Re
|
|
60 kg: 1260 ? ± 100 ? 130 kg: 1260 ? ± 100 ? 280 kg: 1260 ? ± 250 ?
|
1100 ? ± 100 ?
|
13 t: 1200 ? ± 100 ? 28 t: 1075 ? ± 100 ? 60 t: 1350 ? ± 200 ?
|
|
Output resistance Ra
|
|
1020 ? ± 0.5 ?
|
1025 ? ± 25 ?
|
13 t: 1000 ? ± 0.5 ? 28 t: 930 ? ± 0.5 ? 60 t: 1175 ? ± 0.5 ?
|
|
Insulation resistance Ris
|
|
? 20 M?
|
? 5 000 M?
|
? 20 M?
|
|
Rated temperature range Btn
|
|
-10 … +40 °C
|
|
Operating temperature range Btu
|
|
-35 … +70 °C
|
|
Storage temperature range Bts
|
|
-50 … +90 °C
|
|
Sensor material (DIN)
|
|
Stainless steel, mat. no. 14542
|
|
Degree of protection according to EN 60529; IEC 60529
|
|
IP66/68
|
|
Recommended tightening torque of the fixing screws
|
|
8 Nm
|
14 Nm (0.5 … 5 t)
10 Nm (10 t)
|
-
|
|
Current calibration3)
|
|
Standard
|
|
Ex protection to ATEX (optional)
|
|
Available soon
|
|
Cable connection
|
|
|
|
|
|
Function
|
|
Color
|
|
|
|
|
|
pink
|
|
|
|
|
|
gray
|
|
|
|
|
|
brown
|
|
|
|
|
|
white
|
|
|
|
|
|
transparent
|
|
|