The following table lists the measuring conditions for standard applications. The listed values for the measuring range and detection limit (DL) are only approximate values. The exact values at the respective measuring point depend on the totality of all influencing variables and can be determined by Siemens for the specific case. Please note that the values for the detection limit and the maximum measuring range refer to a path length of 1 m. Longer path lengths will improve the detection limit, but not linearly. due to limiting effects such as dust load. The maximum applicable measuring ranges can only be used if permitted by the process conditions such as dust load.
|
Standard application
Effective optical path length: 0.3 … 12 m
Dust load2): < 50 g/Nm3
|
Process gas temperature
Tmin … Tmax
|
Process gas pressure pmin … pmax
|
Min. measuring range
(with 1 m eff. opt. path length)
|
Max. measuring range
(also dependent on eff. opt. path length: see following column)
|
(Max. measuring range x path length)
|
(DL x path length)
|
Accuracy3)
|
| |
|
|
|
|
|
|
|
|
under standard conditions1) without cross-interferences of other gases
|
At 1 013 hPa with cross-interference of gas 2
|
|
|
Gas 1
|
Gas 2
|
Gas code
|
Appl. code
|
|
|
Gas 1
|
Gas 1
|
Gas 1
|
Gas 1
|
Gas 1
|
Gas 1
|
|
O2
|
|
A
|
B6)
|
600 … 1 200 °C
|
950 …1 050 hPa
|
0 … 15 vol%
|
0 … 100 vol%
|
240 vol%*m
|
0.3 vol%*m At 600 °C
|
|
5 %
|
|
C
|
0 … 600 °C
|
950 …1 050 hPa
|
0 … 5 vol%
|
0 … 100 vol%
|
75 vol%*m
|
0.1 vol%*m
|
|
2 %4)
|
|
P
|
0 … 200 °C
|
950 … 5 000 hPa
|
0 … 5 vol%
|
0 … 100 vol%
|
75 vol%*m
|
0.1 vol%*m
|
|
2 %
|
|
NH3
|
|
C
|
A
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 25 ppmv
|
0 … 500 ppmv
|
2 500 ppmv*m
|
0.5 ppmv*m
|
0.9 ppmv*m
At 15 vol% H2O, 55 °C
|
2 %
|
|
T
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 25 ppmv
|
0 … 500 ppmv
|
2 500 ppmv*m
|
0.5 ppmv*m
|
0.9 ppmv*m
At 15 vol% H2O, 55 °C
|
2 %
|
|
E
|
250 … 350 °C
|
950 … 1 050 hPa
|
0 … 45 ppmv
|
0 … 500 ppmv
|
2 500 ppmv*m
|
0.9 ppmv*m
At 250 °C
|
1.4 ppmv*m
At 15 vol% H2O, 250 °C
|
2 %
|
|
F
|
300 … 400 °C
|
950 … 1 050 hPa
|
0 … 50 ppmv
|
0 … 500 ppmv
|
2 500 ppmv*m
|
1 ppmv*m
At 300 °C
|
1.5 ppmv*m
At 15 vol% H2O, 300 °C
|
2 %
|
|
NH3
|
H2O
|
D
|
A
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 25 ppmv
|
0 … 100 ppmv
|
1 200 ppmv*m
|
0.5 ppmv*m
|
0.9 ppmv*m
At 15 vol% H2O, 55 °C
|
2 %
|
|
T
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 25 ppmv
|
0 … 100 ppmv
|
1 200 ppmv*m
|
0.5 ppmv*m
|
0.9 ppmv*m
At 15 vol% H2O, 55 °C
|
2 %
|
|
E
|
250 … 350 °C
|
950 … 1 050 hPa
|
0 … 45 ppmv
|
0 … 100 ppmv
|
1 200 ppmv*m
|
0.9 ppmv*m
At 250 °C
|
1.4 ppmv*m
At 15 vol% H2O, 250 °C
|
2 %
|
|
F
|
300 … 400 °C
|
950 … 1 050 hPa
|
0 … 50 ppmv
|
0 … 100 ppmv
|
1 200 ppmv*m
|
1 ppmv*m
At 300 °C
|
1.5 ppmv*m
At 15 vol% H2O, 300 °C
|
2 %
|
|
HCl
|
|
E
|
A
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 30 ppmv
|
0 … 6 000 ppmv
|
1 200 ppmv*m
|
0.6 ppmv*m
|
2.2 ppmv*m
At 15% H2O, 55 °C
|
5 %
|
|
T
|
120 … 210 °C
|
950 … 1 050 hPa
|
0 … 10 ppmv
|
0 … 60 ppmv
|
720 ppmv*m
|
|
|
|
|
H
|
150 … 250 °C
|
950 … 1 050 hPa
|
0 … 50 ppmv
|
0 … 6 000 ppmv
|
1 200 ppmv*m
|
1.0 ppmv*m
At 150 °C
|
3.1 ppmv*m
At 15 vol% H2O, 150 °C
|
5 %
|
|
HCl
|
H2O
|
F
|
A
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 30 ppmv
|
0 … 100 ppmv
|
1 200 ppmv*m
|
0.6 ppmv*m
|
2.2 ppmv*m
At 15% H2O, 55 °C
|
5 %
|
|
T
|
120 … 210 °C
|
950 … 1 050 hPa
|
0 … 10 ppmv
|
0 … 60 ppmv
|
720 ppmv*m
|
|
|
|
|
H
|
150 … 250 °C
|
950 … 1 050 hPa
|
0 … 50 ppmv
|
0 … 100 ppmv
|
1 200 ppmv*m
|
1.0 ppmv*m
At 150 °C
|
3.1 ppmv*m
At 15 vol% H2O, 150 °C
|
5 %
|
|
HF
|
|
G
|
A
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 5 ppmv
|
0 … 1 500 ppmv
|
200 ppmv*m
|
0.1 ppmv*m
|
0.6 ppmv*m
At 15 vol% H2O, 55 °C
|
5 %
|
|
H
|
150 … 250 °C
|
950 … 1 050 hPa
|
0 … 5 ppmv
|
0 … 1 500 ppmv
|
200 ppmv*m
|
0.11 ppmv*m
At 150 °C
|
0.6 ppmv*m
At 15 vol% H2O, 150 °C
|
5 %
|
|
HF
|
H2O
|
H
|
A
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 5 ppmv
|
0 … 200 ppmv
|
200 ppmv*m
|
0.1 ppmv*m
|
0.6 ppmv*m
At 15 vol% H2O, 55 °C
|
5 %
|
|
H
|
150 … 250 °C
|
950 … 1 050 hPa
|
0 … 5 ppmv
|
0 … 200 ppmv
|
200 ppmv*m
|
0.11 ppmv*m
At 150 °C
|
0.6 ppmv*m
At 15 vol% H2O, 150 °C
|
5 %
|
|
CO
|
|
J
|
C
|
0 … 600 °C
|
950 … 1 050 hPa
|
0 … 1.5 vol%
|
0 … 100 vol%
|
40 vol%*m
|
300 ppmv*m
|
1 500 ppmv*m
at 50 vol% CO2, 20°C
|
2 %
|
|
CO
|
CO2
|
K
|
D
|
0 … 400 °C
|
800 …1 400 hPa
|
0 … 3.0 vol%
|
0 … 100 vol%
|
35 vol%*m
|
0.06 vol%*m
|
0.5 Vol%
at 50 vol% CO2, 20°C
|
2 %5)
|
|
CO2
|
|
L
|
A
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 7.5 vol%
|
0 … 100 vol%
|
40 vol%*m
|
300 ppmv*m
|
|
2 %
|
|
H2O
|
|
M
|
A
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 5 vol%
|
0 … 30 vol%
|
240 vol%*m
|
0.1 vol%*m
|
|
5 %
|
|
T
|
0 … 150 °C
|
950 … 1 050 hPa
|
0 … 5 vol%
|
0 … 30 vol%
|
240 vol%*m
|
0.1 vol%*m
|
5 %
|
1) All technical data apply to an optical path distance of 1 m in a nitrogen atmosphere under standard conditions 25 °C (or Tmin) and 1 013 hPa. The effective detection limit, the measuring range and the accuracy can be influenced by process parameters such as pressure, temperature and gas composition. Not all combinations of maximum pressure and temperature can be realized with the minimum measuring ranges. If the process conditions deviate from the specifications of the standard applications, special applications are also possible on request. Please complete the application questionnaire which can be found on the Internet at www.siemens.de/insitufragebogen.
2) At 0.3 m effective optical path length, average diameter of the dust particles: 15 µm, specific weight of the dust particles: 650 kg/m3
3) At least: Detection limit
4) Up to 200 °C, 5% above this
5) Up to 60 vol. % CO or up to 60 vol. % CO2. Higher CO or CO2 concentrations on request.
6) At high process temperatures, the use of an IR filter A5E00534668 is recommended for the CD 6 sensor (see page 2/xx).
|
Standard application
Effective optical path length: 0.3 … 12 m
Dust load3): < 50 g/Nm3
|
Min. measuring range
(with 1 m eff. opt. path length)
|
Max. measuring range
(usually also dependent on eff. opt. path length: see following column)
|
(Max. measuring range x path length)
|
(DL x path length)
|
Accuracy4)
|
Purging gas mode
|
Purging gas medium
|
| |
|
|
|
|
|
|
under standard conditions1) 2)
|
At 1 013 hPa with cross-interference of gas 1
|
|
|
|
|
|
Gas 1
|
Gas 2
|
Gas code
|
Appl. code
|
Gas 2
|
Gas 2
|
Gas 2
|
Gas 2
|
Gas 2
|
Gas 2
|
Standard
|
Optional
|
|
|
O2
|
|
A
|
B6)
|
|
|
|
|
|
|
E, F
|
G, H
|
Steam + air, N2
|
|
C
|
|
|
|
|
|
|
D
|
B
|
N2
|
|
P
|
|
|
|
|
|
|
D
|
B
|
N2
|
|
NH3
|
|
C
|
A
|
|
|
|
|
|
|
C
|
G
|
Air
|
|
T
|
|
|
|
|
|
|
C
|
G
|
Air
|
|
E
|
|
|
|
|
|
|
E
|
G
|
Air
|
|
F
|
|
|
|
|
|
|
E
|
G
|
Air
|
|
NH3
|
H2O
|
D
|
A
|
0 … 5 vol%
|
0 … 30 vol%
|
240 vol%*m
|
0.1 vol%*m
|
0.1 vol%*m
|
5 %
|
C
|
G
|
Air
|
|
T
|
0 … 5 vol%
|
0 … 30 vol%
|
240 vol%*m
|
0.1 vol%*m
|
0.1 vol%*m
|
5 %
|
C
|
G
|
Air
|
|
E
|
0 … 5 vol%
|
0 … 30 vol%
|
240 vol%*m
|
0.1 vol%*m
At 250 °C
|
0.1 vol%*m
At 250 °C
|
5 %
|
E
|
G
|
Air
|
|
F
|
0 … 5 vol%
|
0 … 30 vol%
|
240 vol%*m
|
0.1 vol%*m
At 300 °C"
|
0.1 vol%*m
At 300 °C"
|
5 %
|
E
|
G
|
Air
|
|
HCl
|
|
E
|
A
|
|
|
|
|
|
|
C
|
G
|
Air
|
|
T
|
|
|
|
|
|
|
C
|
G
|
Air
|
|
H
|
|
|
|
|
|
|
E
|
G
|
Air
|
|
HCl
|
H2O
|
F
|
A
|
0 … 5 vol%
|
0 … 30 vol%
|
360 vol%*m
|
0.1 vol%*m
|
0.1 vol%*m
|
5 %
|
C
|
G
|
Air
|
|
T
|
0 … 5 vol%
|
0 … 30 vol%
|
360 vol%*m
|
|
|
|
C
|
G
|
Air
|
|
H
|
0 … 5 vol%
|
0 … 30 vol%
|
360 vol%*m
|
0.1 vol%*m
At 150 °C
|
0.1 vol%*m
At 150 °C
|
5 %
|
E
|
G
|
Air
|
|
HF
|
|
G
|
A
|
|
|
|
|
|
|
C
|
G
|
Air
|
|
H
|
|
|
|
|
|
|
E
|
G
|
Air
|
|
HF
|
H2O
|
H
|
A
|
0 … 5 vol%
|
0 … 30 vol%
|
360 vol%*m
|
0.1 vol%*m
|
0.1 vol%*m
|
5 %
|
C
|
G
|
Air
|
|
H
|
0 … 5 vol%
|
0 … 30 vol%
|
360 vol%*m
|
300 ppmv*m
At 200 °C
|
300 ppmv*m
At 200 °C
|
5 %
|
E
|
G
|
Air
|
|
CO
|
|
J
|
C
|
|
|
|
|
|
|
E
|
G
|
Air, N2
|
|
CO
|
CO2
|
K
|
D
|
0 … 7.5 vol%
|
0 … 100 vol%
|
75 vol%*m
|
0.15 vol%*m
|
0.5 vol% at 50 vol% CO, 20 °C
|
2 … 5 %5)
|
C
|
G
|
Air
|
|
CO2
|
|
L
|
A
|
|
|
|
|
|
|
C
|
G
|
Air
|
|
H2O
|
|
M
|
A
|
|
|
|
|
|
|
C
|
G
|
Air
|
|
T
|
|
|
|
|
|
|
C
|
G
|
Air
|
1) At 20 °C, 1 013 hPa
2) If the smallest permissible process gas temperature of the application is Tmin > 20 °C, the detection limit refers to Tmin and standard pressure (1 013 hPa)
3) At 0.3 m optical path length, average diameter of the dust particles: 15 µm, specific weight of the dust particles: 650 kg/m3
4) At least: Detection limit
5) Depends on temperature (higher values at higher temperatures)
6) At high process temperatures, the use of an IR filter A5E00534668 is recommended for the CD 6 sensor (see page 2/xx).
Special applications
| |
|
 |
If the process conditions deviate from the specifications of the standard applications, special applications are also possible on request. Please complete the application questionnaire which can be found on the Internet at www.siemens.de/insitufragebogen.
|
|
|